TESTING
TESTING
6
7
®
1. Remove the endcap
from the lancing device.
Hold the endcap dial
and gently snap off
the endcap from top
to bottom.
WARNING: Potential Biohazard
The lancing device provided by Bayer is intended for
single patient use only. The lancing device should
not be used for assisted blood draws by health
care providers or at health care provision sites, and
should never be shared with anyone else, even a
family member, due to risk of infection. Use a new
lancet each time you test because it is no longer
sterile after use.
Preparing the Lancing Device
Getting Ready to Test
Getting Ready to Test
WARNING
Always wash your hands well with soap and water
and dry them well before and after testing, handling
the meter, lancing device or used test strips.
Read your CONTOUR
®
NEXT user guide, the lancing
device insert and all instructional materials provided
in your meter kit before testing. Please follow all
instructions for use and care exactly as described.
Examine product for missing, damaged, or broken
parts. If the test strip bottle is open inside a new box
of strips, do not use those strips. Contact Bayer
Diabetes Care Helpline at 1-800-268-7200 for
replacement parts.
CAUTION: Your C
ONTOUR
®
NEXT meter only works
with C
ONTOUR
®
NEXT test strips and CONTOUR
®
NEXT
control solutions.
Have all the materials you will need ready before
you begin testing. This includes your C
ONTOURNEXT
meter, C
ONTOUR NEXT test strips, and the lancing
device and lancets from Bayer. You may also need
C
ONTOURNEXTcontrol solution to perform a quality
control check. C
ONTOUR NEXT test strips and
C
ONTOURNEXT control solutions are sold separately
if not included in the meter kit.
2. Loosen the round
protective cap on a
lancet by rotating it
¼turn, but do not
remove it.
Preparing the Lancing Device
The lancing device graphics are for illustration purposes
only. Your lancing device may look different. Refer to
your lancing device insert for detailed instructions on
preparing the lancing device.
For Alternative Site Testing instructions, see page 16.
WARNING
Dispose of the used lancet as medical waste or as
advised by your health care professional.
Do not reuse lancets. Use a new lancet each
time you test.